Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 86: CAL ADJ /5500 INSTRUMENT: Fluke 88: CAL ADJ /5500 DATE: 23-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 77 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 86,88 Series service manual (PN 880906 December 1990). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 80_pca.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P J F W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP Disconnect the test leads if attached. Remove the three 1.003 DISP phillips-head screws from the bottom of the case. Turn 1.003 DISP the instrument face-up, and grasp the top cover at both 1.003 DISP sides of the input connectors. Then, pull the top cover 1.003 DISP from the UUT. 1.004 DISP Turn the UUT on. 1.004 DISP Ambient temperature: 18C - 28C. 1.004 DISP Relative humidity: less than 80%. 1.004 DISP Allow the UUT to stabilize. 1.005 HEAD DC VOLTAGE CALIBRATION 1.006 DISP Select the V DC function on the UUT. 1.006 DISP 1.006 DISP Press RANGE for 2s to select autorange. 1.007 JMPT 1.010 Fluke 88: CAL ADJ /5500 1.008 PIC 55_86_2W 1.009 JMP 1.011 1.010 PIC 55_88_2W # 1.011 DISP Connect the 5500A and the UUT as follows. # 1.011 DISP [32] 5500A NORMAL HI to UUT V Ohm Diode # 1.011 DISP [32] 5500A NORMAL LO to UUT COM 1.011 5500 3.5V S 2W 1.012 PIC 80_PCA (20,3,R/W) "Adjust R21 for a UUT reading of" 1.012 PIC (22,3,R/W) "3.500+/-0.001V" 1.013 5500 * S 1.014 HEAD AC VOLTAGE CALIBRATION 1.015 DISP Select the V AC function on the UUT. 1.015 DISP 1.015 DISP Press RANGE for 2s to select autorange. 1.016 5500 3.5V 100H SI S 2W 1.017 PIC 80_PCA (20,3,R/W) "Adjust R34 for a UUT reading of" 1.017 PIC (22,3,R/W) "3.500+/-0.002V" 1.018 5500 3.5V 10kH SI S 2W 1.019 PIC 80_PCA (20,3,R/W) "Adjust C2 for a UUT reading of" 1.019 PIC (22,3,R/W) "3.500+/-0.010V" 1.020 5500 35V 10kH SI S 2W 1.021 PIC 80_PCA (20,3,R/W) "Adjust C3 for a UUT reading of" 1.021 PIC (22,3,R/W) "35.00+/-0.10V" 1.022 5500 * S 1.023 HEAD 1.024 END